SPERM CRYO SYSTEM 0.25 ml
IVF CONSUMABLES

SCS-620


SPERM CRYO SYSTEM, 0.25 ml, Transparent,

For Sperm Cryopreservation


A. USAGE:

  • SPERM CRYO SYSTEM are specialized tool used for sperm cryopreservation in Assisted Reproductive Technologies (ART).


B. STORAGE:

  • Store in a clean, dry environment, away from light and heat sources.


C. STERILITY:

  • Sterile A


D. DESCRIPTION:

  • Straw length 133 mm, flexible, transparent, made of high-quality materials.
  • This straw is composed of two distinct sections separated by a white safety plug.


E. PRECAUTIONS:

Handling and storage:

  • This product is for laboratory use only and should be handled by trained professionals.
  • Always handle straws with cryo-safe gloves to avoid thermal injury.
  • Store straws in labeled canisters within the liquid nitrogen tank to prevent misidentification.

Avoid overfilling:

  • Overfilling straws can lead to cracking or rupture because the sample volume increase during freezing process that make excessive pressure.
  • To prevent these issues, always leave a small air gap inside the straw to accommodate expansion, and begin freezing in nitrogen vapor (-80 °C to -120 °C) for gradual cooling before immersing in liquid nitrogen (-196 °C).

Sterility:

  • Single use only.
  • Avoid touching the open ends of the straw to avoid contamination.

Freezing process:

  • Always use a gradual cooling process to prevent thermal shock and ensure sample viability.

Labeling:

  • Use permanent, freezing resistant labels to prevent sample data under freezing conditions.

Equipment safety:

  • Use appropriate safety equipment when working with liquid nitrogen to avoid burns.


F. PROCEDURES:

Preparation:

  • Collect and evaluate the sample for quality, adjusting it with a cryoprotective freezing medium to protect sperm cells during freezing.
  • Verify that the straws are sterile and the package is intact before use.
  • Ensure all tools and cryoprotective agents are pre-prepared and at the required temperatures.
  • Maintain an aseptic environment to avoid contamination.


Filling & Sealing:

  • Fill straws with sample and seal them, ensuring proper labeling with details and freezing date.
  • Load the sample (e.g., semen sample) using a filling nozzle and a micro-aspirator or a syringe with a specific fitting.
  • Do not fill the straw completely, leave a small air gap inside the straw (about 10 % of the volume) to accommodate expansion during freezing.
  • The sample should rise through the white safety plug up to half of the powder.
  • Seal both ends of the straw securely using a heat sealer.
  • Label each straw clearly with sample ID, date, and any other necessary information.


Freezing:

  • Place straws horizontally in a controlled freezing device or in nitrogen vapor (-80 °C to -120 °C) for
  • 15 – 30 minutes.
  • Transfer the straws into liquid nitrogen tank (-196 °C) for long term storage.


Thawing:

  • Retrieve the straw carefully from the liquid nitrogen using cryo-safe gloves and tools.
  • Place the straw in a water bath at the appropriate thawing temperature (e.g., 37 °C) for a few seconds.

  • Quantity : 100 PCS
  • Storage : Room Temperature
  • Cat. No. : SCS-620
  • Sterility : Sterile A

Download SPERM CRYO SYSTEM 0.25 ml PRODUCT INSERT.pdf